6-3
Routine Maintenance Operator Manual 122994-513
4.0 CHAMBER COMPONENTS (Cont’d)
4.5 Verify proper water level and temperature.* 3 X/year
4.6 Verify piping system for leaks. Repair if necessary.* 6 X/year
4.7 Verify condition of sliding guides. Replace if necessary.* 3 X/year
5.0 SUPPLY-LINE STRAINERS
Inspect strainers for debris. Clean as necessary (see S
ECTION 6.5.2). 6 X/year
6.0 CHEMICALS AND INJECTION PUMPS
6.1 Verify suction tube is not clogged, cracked or distorted. If damaged, call STERIS.* 6 X/year
6.2 Inspect and lubricate squeeze tubes. Replace if necessary (see S
ECTION 6.7). 6 X/year
6.3 Verify amount of chemical product injected. Adjust if necessary. 3 X/year
6.4 Lubricate pump roller (see S
ECTION 6.7). 6 X/year
7.0 COMPRESSOR
Verify intake filter and mufflers. Change if necessary.* 2 X/year
8.0 PROTECTIVE GROUND TERMINAL*
Verify protective conductor at equipment terminal and verify if connection is well
secured inside terminal with proper torque equipment. Torque requirement for
supply conductor terminals for steam-heated units (L1-L2-L3- N) is: 7.1 - 8.9 lb/in
(0.8 -1.0 N•m) and for electric-heated units (L1-L2-L3- N) is: 7.1 - 8.9 lb/in (0.8 -
1.0 N•m); for steam- and electric-heated units, protective conductor terminal is:
20 lb/in (2.2 N•m).*
1 X/year
9.0 FINAL TEST
9.1
Clean dirt and lint from components. Verify all wiring, terminals and socket connec-
tions for damage or fraying.
6 X/year
9.2 Verify glassware washer has proper labels (caution, warning, etc.). 6 X/year
9.3 Install any panel or cover removed during inspection. 6 X/year
9.4 Test glassware washer operation. 6 X/year
*Contact STERIS for this service. Preventive Maintenance is not covered under warranty.
Table 6-1. Preventive Maintenance Guide (Cont’d)
Recommended frequency of inspection is indicated in the right column. Usage and
utility conditions may require more or less frequent inspections.
Suggested
Frequency