Controls, Connectors, and Indicators-2465 Operators
XI0
MAG
Switch-Horizontally magnifies the por-
tion of the sweep display positioned at the center
vertical graticule line by
a
factor of 10 when pressed
in. When the A trace and the B trace are displayed
alternately (TUR N-ALT Horizontal Display Mode
selected), only the B trace
is
magnified. Using XI0
magnification extends the fastest sweep speed to
500 ps per division. The push button must be pressed
in
a
second time to release
it
and regain the XI sweep-
speed magnification.
AV Switch-Activates the Delta Volts (AV) measure-
ment function, when momentarily pressed in alone,
and cancels any other Delta measurement function
in effect.
In the A Sweep mode (A and
B
SECIDIV
switches locked together), two horizontal cursors
are superimposed on the crt display. The crt readout
displays the equivalent voltage represented by the
separation between the two cursors. The position of
one cursor on the display
is
set by the A REF OR
DLY POS control and the position of the other
is
set
by the A control. With multiple-channel displays,
the deflection
factor of the first channel selected in
the display sequence determines the scale factor of
the Delta Volts readout on the crt. The Delta Volts
readout
is
displayed
as
a percentage ratio if either one
of the following conditions exists:
(I)
the channel
determining the scale factor
is
uncalibrated (VAR
control out of detent), or (2) ADD is displayed alone
when the Channel 1 and Channel 2 deflection factors
are not the same (VOLTSIDIV switches are
at
dif-
ferent settings or are uncalibrated). Either pressing in
the AV switch or pulling the B SECIDIV knob to
the out position when the Delta Volts function
is
active, cancels
it.
Attempting to activate the Delta
Volts function while the A and B SECIDIV knobs
are unlocked causes the message NO AV
WITH
DELAY
to appear in the top row of the crt readout.
If displayed, the error message will be canceled
(removed from the display) by any of the following
actions: pressing either the AV or At switch; pushing
in the
B
SECIDIV if it
is
out or pulling
it
out if it is
in; or locking the A and B SECIDIV knobs together
(set to the same sweep speed with the
B
SECIDIV
knob in).
At Switch-Activates the Delta Time measurement
function and cancels any other Delta measurements
in effect, when momentarily pressed in alone. When
the Delta Time function
is
active, momentarily
pressing in the At push button cancels the function.
When the A and
B
SECIDIV knobs are locked together
(A trace only), two vertical cursors are superimposed
on the crt display while the Delta Time function
is
active. In any of the delay-time Horizontal Display
modes (PULL-INTEN, TURN-ALT, or PUSH-B),
two separate delay times are established by the
Delta Time function. One cursor position (or delay
time)
is
set
by the A REF OR DLY POS control,
and the other
is
set
by the A control. The crt readout
displays either the time difference between the two
delays or the equivalent time difference between the
two vertical cursors.
If the SECIDIV VAR control
is
not in the detent
position, At cursor difference on the A trace only
displays
is
expressed
as
a
ratio, with five divisions
corresponding to
a
100% ratio. For the delay-time
Horizontal Display modes, the SECIDIV VAR
control varies the B-sweep scale factor
as
it
is
rotated,
but
it
has no effect
on
the delay time readout.
Pressing in the AV and At push buttons together
activates the
l/At measurement function and cancels
any other Delta measurement functions in effect. The
crt waveform display and operation of both the
A
REF OR DLY POS and A controls remain the same
as
explained for At operation. However, with l/At
selected, the crt readout shows the reciprocal of the
time-difference measurement, with units being
frequency (Hz, kHz, MHz, or GHz).
For A trace only displays, with the SECIDIV VAR
control out of the detent (fully clockwise) position,
the time difference between
I
/At cursors
is
displayed
in degrees of phase, with five divisions equal to 360
degrees. As with At measurements, the position of
the SECIDIV VAR control has no effect on delay-
time displays except to change the B Sweep scale
factor, and the readout remains in units of frequency.
When the l/At function
is
active, pressing both the
AV and the At push buttons together cancels the
function and exits the Delta measurement mode.
Pressing either AV or At alone cancels the 1IAt
function and activates the function associated with
the button pressed.
A
REF OR DLY
POS Control-Sets the reference
B
Sweep delay time or positions the Reference cursor
when AV, At, or l/At Measurement Mode is active.
When any delay-time Horizontal Display Mode
(PU
LL-INTEN, TURN-ALT, or PUSH-B)
is
selected,
the reference
B
Sweep delay time is determined by
the rotation of the A REF OR DLY POS control in
conjunction with the A SECID IV switch setting.
This control produces fine resolution over
a
portion
of
its
range, after which it changes to coarse reso-
lution. It reenters the fine-resolution range upon
reversing the direction of rotation.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2465 and is the answer not in the manual?

Tektronix 2465 Specifications

General IconGeneral
BrandTektronix
Model2465
CategoryTest Equipment
LanguageEnglish

Related product manuals