air
force
T033A1-13-496-1
NAVELEX
0969-LP-170-0010
Special
Tools
and
Test
Equipment
SECTION
II
SPECIAL
TOOLS
AND
TEST
EQUIPMENT
2-1.
SPECIAL
TOOLS.
No
special
tools
are
required.
2-2. TEST
EQUIPMENT.
Test
equipment
required
to
maintain
the
instrument
is
listed
in
Table
2-1.
Equiv-
alent
items
may
be
used
if the
recommended
items
are
not
available.
2-3.
CONSUMABLE
MATERIALS.
Table
2-2
lists
the
consumable
materials
recommended
for
maintaining
the
instrument.
Equivalent
materials
may
be
used
if those
re-
commended
are
not
available.
Table
2-1. Test
Equipment
List
Tool/Equipment
Number
Nomenclature
Application
Description
Tektronix
PG
506^
Calibration
Generator
and
Fast-rise
Pulse
Generator
Vertical
deflection
system
checks and
adjustments:
trigger
range
check;
trigger
view
check
and
adjustment;
high
and
low
frequency
com-
pensation
adjustments.
Range, 1
kilohertz
and 100
kilo-
hertz
square-wave;
output
ampli-
tude,
20
millivolts
to 20
volts; ac-
curacy,
within 0.4%;
fast-rise out-
put
risetime,
1
nanosecond or
less.
Tektronix
TG
501*
Time-Mark
Generator
Sweep
timing
checks
and ad-
justments;
Y-axis
adjust-
ments;
geometry
adjust-
ments.
Marker
range, 10
nanoseconds
to
0.5
seconds;
accuracy,
within
0.4%.
Tektronix
SG
502*
Sine-wave
Generator
Low
Frequency
Trigger
checks.
Range, 30
hertz
to
50
kilohertz;
output
amplitude,
10
millivolts to
4 volts
peak to
peak.
Tektronix
SG
503*
Sine-wave
Generator
Bandwidth
checks;
cascade
sensitivity
checks;
trigger
checks
and
adjustments.
Range, 4
megahertz to 100
mega-
hertz with a
50
kilohertz
refer-
ence;
accuracy,
within
3%;
output
amplitude, 5
millivolts to
4 volts
peak
to
peak.
Tektronix
DM
501*
Digital
Multimeter
Power
supply
checks
and ad-
justments;
calibrator
adjust-
ments; crt
bias
adjustments;
troubleshooting.
Range,
—10 volts dc to
+50
volts
dc; 300
volts
ac, 2
kilohm to
20
megohm;
accuracy,
within 0.1%.
AN/USM425(V)1
Oscilloscope
Sweep
gate
output
checks;
cali-
brator
output
checks;
Z-axis
compensation
adjustment
and
calibration
checks;
trouble-
shooting
waveforms.
Bandwidth,
at
least
100
megahertz;
vertical
deflection
factor, at
least
5
millivolts/division;
sweep
rate,
at
least 2
microseconds/division.
Tektronix
part
017-0061-00
CT-3
Signal
Pickoff
Trigger
checks.
Assembly,
signal
pickoff
(CT-3).
2-1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 465M and is the answer not in the manual?

Tektronix 465M Specifications

General IconGeneral
BrandTektronix
Model465M
CategoryTest Equipment
LanguageEnglish

Related product manuals