(5)
Adjust XI 0
GAIN, R2925
for one
time mark
per
division.
(6)
Continue to
the next
procedure.
q.
Magnifier
Registration
(Figure
5-22).
(1)
SetXIOMAG to
In (on).
(2)
Adjust
horizontal
POSITION
until the
sweep
starts at
the
center
vertical
graticule line.
(3)
Set X10
MAG
to
Out
(off).
(4)
Adjust
MAG
REG, R2932
until the
sweep starts
at
the
center
vertical
graticule line.
INTERACTION
NOTE
R2932
and
horizontal POSITION
may interact;
therefore, repeat
steps
(1)
through
(4)
until
no
further
adjustment of R2932 is
needed.
AIR
FORCE T033A1-13-496-1
NAVELEX
0969-LP-170-0010
Maintenance Instructions
(5)
Continue to
the next
procedure.
r.
B
Sweep
Timing
(Figure
5-23).
(1
)
Set controls as
follows:
XI
0
MAG
Out (off)
A
AND
BTIME/DIV
1ms
HORIZ
DISPLAY
B DLY'D
(2)
Set time mark
generator for one millisecond
time marks.
(3)
Set
horizontal POSITION to
align the first
time
mark with the
left vertical graticule line.
(4)
Adjust R2682
for one time mark per division.
»
(5)
Continue to
the next procedure.
Figure
5-23.
Sweep
adjustment
locations.