AIR
FORCE
T033A1
-13-496-1
NAVELEX
0969-LP-170-0010
Maintenance
Instructions
Table
5-1.
Operational Checkout
Procedures—Continued
Step
Procedure
Normal Indication
b.
Set the
LINE
RANGE
Selector
for the correct
power source
voltage range
and check the
fuse size
as
indicated on the
instrument rear
panel.
Check that
the POWER switch is
off.
ON
indicator is not lit.
d.
Connect
the power cord to the
instrument and the
power source.
e.
Pull on the
POWER switch. Let the
instrument warm up
for
5
minutes,
if it was
stored
at or
above
a 0°C
(-^32°F) ambient temperature;
other-
wise, 20
minutes.
ON
indicator is lit.
NOTE
For the best
overall performance of the
instrument, the
checkout pro-
cedures should
be made within an
ambient temperature
range
of
+20°
to
+30°
C
(+68°
to
+86°
Fj.
2
.
Preliminary Adjustments.
Adjust
INTEN for
a
low intensity trace.
Adjustable
from off
to
blooming.
b.
Set
CH
1
AC-GND-DC
to
GND. Adjust
FOCUS for a
well
defined trace.
Adjustable on either
side of an opt-
imum setting.
c.
Position
the trace
to the center
horizontal graticule line.
3. Vertical
Deflection Factor
Accuracy.
a. Connect the equipment as follows:
Trace is parallel
to
the horizontal
graticule line.
PG
506
OR
EQUIVALENT
CALIBRATION
GENERATOR
5-2