Installing Your
Instrument
Operating Requirements
CAUTION. To en
sure proper cooling, keep sides of the instrument clear of obstructions.
Place the instrument on a cart or bench, observing clearance requirements:
Top: 20 mm (0.8 in)
Left and right side: 150 mm (5.9 in)
Bottom: 20 mm (0.8 in)
Rear: 75 mm (3 in)
Environmental Req
uirements
Before operating the instrument, ensure that the instrument has warmed up for 20 minutes and meets the environmental
requirements lis
ted in the following table.
Requirement Description
Temperature (op
erating)
10 °C to 40 °C (+50
°F to +104 °F)
Relative humid
ity (operating)
5% to 95% relati
ve humidity at up to 30 °C (86 °F)
5% to 45% relative humidity above 30 °C (86 °F) up to
+40 °C (104 °F) noncondensing, and as limited by a
maximum wet-bu
lb temperature of 29 °C (84.2 °F)
Altitude (ope
rating)
Up to 3,000 m (a
pproximately 10,000 feet)
Power Supply Requirements
WARNING. To r
educe the risk of fire and shock, ensure that the mains supply voltage fluctuations do not exceed 10% of the
operating voltage range.
Source Voltage and
Frequency
Power Consumption
100 VAC to 240 VAC,
47 Hz to 63 Hz
Less than 450 W (AW
G7000 series)
Less than 560 W (AWG5000 series)
2 AWG5000 and AWG7000 Series Quick Start User Manual