Adjustment Procedures
d. Measure R1 to C1
Delay using the Measurement function of the sampling
oscilloscope. Select the following parameters in the Inter-Channel Skew
Calibration window:
SamplingRate: 6.0GHz
Ch 2 Phase: 0 °
e. If the delay value measured in step d is outside ±83 ps range, change
the Sampling Rate to 1 2.0 G Hz and then return the setting to 6.0 GHz.
Measure R1 to C1 Delay again. Repeat this measurement until the R1
to C1 Delay value is within the ± 83 ps range.
f. Change the parameters as follows in the Inter-Channel Skew Calibration
window:
SamplingRate: 6.0GHz
Ch 2 Phase: +180 °
g. Enter the measurement values of R1 to C1 Delay for each setting to the
rst place after the decimal point while changing the Ch 2 Phase settings
from +180 degrees to -180 degrees using either the mouse or the Tab key.
Adjust the horizontal position of the sampling oscilloscope as necessary
so that the rising edge of the waveform is displayed at the center of the
screen.
h. After you complete all the measurements for SamplingRate:6.0GHz,
click Calc (Delay to Phase). This will display the calculated phase values
in the Ch1– Ch2 Phase (degree) eld.
i. Select Sampling Rate: 12.0 GHz.
j. Disconnect the SMA cable from the Ch 2 Analog (+) Output, and then
connectittotheCh1Analog(+)Output.
k. Adjust the horizontal position of the sampling oscilloscope so that the
rising edge of the arbitrary waveform generator Ch 1 Analog (+) Output
waveform is visible near the center of the oscilloscope screen, and
then
select File Menu > Save Waveform... to save the waveform data to R1.
l. Disconnect the SMA cable from the Ch 1 Analog (+) Output, and then
connectittotheCh2Analog(+)Output.
m. Change the parameters as follows in the Inter-Channel Skew Calibration
window:
Sampling Rate: 12.0 GHz
Ch 2 Phase: +180 degree
n. Enter the measurement values of R1 to C1 Delay for each setting to the
rst place after the decimal point while changing the Ch 2 Phase settings
in the o rder from +180 degrees to -180 degrees using either the mouse or
the Tab key. Adjust the horizontal position of the sampling oscilloscope
3–10 AWG7000B and AWG7000C Series Service Manual

Table of Contents

Other manuals for Tektronix AWG7082C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix AWG7082C and is the answer not in the manual?

Tektronix AWG7082C Specifications

General IconGeneral
BrandTektronix
ModelAWG7082C
CategoryPortable Generator
LanguageEnglish

Related product manuals