Trigger Setup
Trigger Coupling
Trigger coupling determines what part of the
signal is passed to the trigger circuit. Edge
triggering can use all available cou pling types:
DC, AC, Low Frequency Rejection, High
Frequency Rejection, and Noise Rejection.
Sequence (B) triggering does not provide AC
coupling. Other trigger types use DC coupling
only.
Horizontal Position
When Delay Mode is on, use horizontal
position to acquire waveform detail in a region
that is separated from the trigger location by a
signicant interval of time.
110 DPO3000 Series Oscilloscopes User Ma nual