Get Acquainted with the Instrume nt
Rear-Panel Connectors
1. Cal. This is for use by authorized service
personnel only.
2. Aux Ou t. Use this output to synchronize
other test equipment with your oscilloscope.
A LOW to HIGH transition indicates that the
trigger occu rred. The logic level for Vout
(HI) is ≥3.25V open circuit; ≥2.2 V into a
50Ω load to ground. The logic level for
Vout (LO) is ≤0.4Vintoaloadof≤4mA;
≤0.2Vintoa50Ω load to ground.
3. Probe Power. Use this port to provide
auxiliary power for TekVPI probes, when
needed.
4. Video Out. Use the Video Out port ( DB-15
female connector) to show the oscilloscope
display on an external m onitor or projector.
5. LAN. Use the LAN (Ethernet) port (RJ-45
connector) to connect the oscilloscope to a
10/100 Base-T local area network.
74 DPO3000 Series Oscilloscopes User Ma nual