Installation
Aux In: 1 M
The maximum input voltage: At the front-panel connector, 300 V
RMS
, Installation Category II; for measurements performed on
circuits directly connected to the low-voltage installation.
CAUTION. To ensure proper cooling, keep the sides and rear of t he instrument clear of obstructions.
Total Probe Power:
If the total probe power requirements exceed the available power from the oscilloscope, connect the external AC adapter (Tektronix
part number 119-7465-XX) to the rear-panel P robe Power connector.
Maximum Probe Powe r Available Per Channel (3 or 5 TekVPI Interfaces):
5 V ± 5%, 50 mA max., 250 mW max.
12 V ± 10%, 2 A max., 24 W max.
10 DPO3000 Series Oscilloscopes User Ma nual