This is the propagation path skew and ignores skew contributions due to threshold inaccuracies
(see Threshold accuracy) and sample binning (see Digital channel timing resolution). Factory
calibration/deskew is required to achieve this number.
Display specificationsContainer
Display type
Display area: 210.4 mm (8.28 in) (H) x 157.8 mm (6.21 in) (V), 264 mm (10.4 in) diagonal, 6-bit
RGB full color, XGA (1024 x 768) TFT liquid crystal display (LCD).
Display resolution 1024 X 768 XGA display resolution
Luminance, typical 400 cd/m
2
Waveform display color scale The TFT display can support up to 262,144 colors. A subset of these colors are used for the
oscilloscope display, all of which are fixed colors and not changeable by the user.
Input-Output port specifications
Ethernet interface Standard on all models: 10/100/1000 Mbps
GPIB interface Available as an optional accessory that connects to the USB Device and USB Host ports with the
TEK-USB-488 GPIB to USB Adapter
The control interface is incorporated into the instrument user interface.
Video signal output A 15 pin D-sub, VGA-type connector, supporting a minimum resolution range of 640X480 to
1280X1024.
USB interface 1 USB device connector (2.0 Full Speed), on the instrument rear panel.
6 USB host connectors (2.0 HS), two on the instrument front and four on the rear.
Probe compensator output voltage
and frequency, typical
Output voltage 0 V to 2.5 V ±1% behind 1 kΩ ±2%
Frequency 1 kHz ±25%
Auxiliary output (AUX OUT), trigger
out or reference clock out
You can set the Auxiliary output to Trigger Out or Reference Clock Out.
Reference Clock Out: Outputs the 10 MHz oscilloscope reference clock.
Trigger Out: A HIGH to LOW transition indicates that the trigger occurred.
Specifications (MSO/DPO5000/B Series) P6616 digital probe specifications (cont.)
94 MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series

Table of Contents

Other manuals for Tektronix DPO5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO5000 Series and is the answer not in the manual?

Tektronix DPO5000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals