Variable A event trigger holdoff
range
250 ns to 12 s + random holdoff
Lowest frequency for successful
set level to 50%, typical
50 Hz
Trigger jitter (RMS)
≥4 GHz models 100 fs using enhanced Trigger placement. 1.0 ps RMS (typical) for low frequency, fast rise time
signal, A edge, time holdoff = 30 μs.
<4 GHz models 1.5 ps RMS (typical)
Serial trigger specifications
Low speed serial trigger, ≥4 GHz
models
Number of bits 128 bits
Baud rate limits 10 Mbits/s
High speed serial trigger number
of bits
≥4 GHz models 64 bits for NRZ data rates ≤ 1.25 GBaud
40 bits for 8b/10b data rates between 1.25 GBd and 6.25 GBd
<4 GHz models 64 bits
Trigger specifications (cont.) Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series 65