Trigger specifications
Trigger jitter DC coupled A edge
random holdoff, typical
≥4 GHz models 100 fs using enhanced trigger placement.
1 ps rms for low frequency, fast rise time signal, A edge, time holdoff = 30 μs
<4 GHz models 1.5 ps rms for low frequency, fast rise time signal
(DPO7104C and DPO7054C: 3 ps total trigger jitter.)
Edge trigger sensitivity DC
coupled, typical
All sources, positive or negative edge, for vertical scale settings ≥10 mV/div and ≤1 V/div
≥4 GHz models
Trigger Source Sensitivity
A Event trigger ≤ 4%FS from DC to 50 MHz
≤ 10%FS at 4 GHz
≤ 15%FS at 6 GHz
≤ 20%FS at 8 GHz
≤ 50%FS at 11 GHz
20%FS from DC to instrument bandwidth
(typical, enhanced trigger on, auto mode only for
frequencies above 8 GHz)
B Event trigger ≤ 4%FS from DC to 50 MHz
≤ 10%FS at 4 GHz
≤ 15%FS at 6 GHz
≤ 50%FS at 9 GHz
20%FS from DC to instrument bandwidth
(typical, enhanced trigger on, auto mode only for
frequencies above 6 GHz)
Auxiliary input 250 mV from DC to 50 MHz, increasing to
350 mV at 1 GHz
<4 GHz models
Trigger Source Sensitivity
Main and Delayed trigger ≤ 0.7 div from DC to 50 MHz
≤ 1.2 div at 2.5 GHz
≤ 2.5 div at 3.5 GHz with BWE on
Auxiliary input 1 MΩ: 250 mV from DC to 50 MHz, increasing to
350 mV at 250 MHz
50 Ω: 150 mV from DC to 50 MHz, increasing to
200 mV at 1.8 GHz
Video trigger, Ch1 - Ch4 0.6 to 2.5 divisions of video sync tip
Trigger sensitivity TPC, typical The minimum signal levels required for stable triggering of a waveform when the hardware is
incapable of doing so (excessive frequency, small input).
>20 GHz models
Trigger Source Sensitivity
A Event trigger, Ch1 – Ch4 20%×FS from DC to instrument bandwidth
B Event trigger, Ch1 – Ch4 20%×FS from DC to instrument bandwidth
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series)
58 MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series