Sine wave generator leveling procedure
Some procedures in this manual require a sine wave generator to produce the
necessary test signals. If you do not have a leveled sine wave generator, use one
of the following methods to level the output amplitude of your sine wave
generator.
Method 1
Equipment required Prerequisites
Sine wave generator (Item 9)
Meter, power and sensor (Item 10)
Power splitter (Item 11)
50 Ω precision cable 2.92 mm male-to-female
(Item 12)
One K male-to-male adapter (Item 13)
(See Prerequisites on page 116.)
Figure 46: Sine wave generator leveling equipment setup
1. Install the test hookup: Connect the equipment as shown in the above figure.
2. Set the Generator:
Set the sine wave generator to a reference frequency of 10 MHz.
Adjust the sine wave generator amplitude to the required number of
divisions as measured by the instrument.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series 281

Table of Contents

Other manuals for Tektronix DPO5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO5000 Series and is the answer not in the manual?

Tektronix DPO5000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals