EasyManuals Logo
Home>Tektronix>Test Equipment>DPO5000 Series

Tektronix DPO5000 Series Technical Reference

Tektronix DPO5000 Series
352 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #289 background imageLoading...
Page #289 background image
Check serial trigger clock recovery range <4 GHz models
Equipment required Equipment required
One precision 50 Ω coaxial cable (Item 4)
One sine-wave generator (Item 9)
One adapter (Item 19)
The oscilloscope must meet the prerequisites.
Also, the instrument must have passed Check
DC Voltage Measurement Accuracy. (See
Signal acquisition system checks on page 176.)
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series 277

Table of Contents

Other manuals for Tektronix DPO5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO5000 Series and is the answer not in the manual?

Tektronix DPO5000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals