Prerequisites
The tests in this section comprise an extensive, valid confirmation of
performance and functionality when the following requirements are met:
The cabinet must be installed on the instrument.
You must have performed and passed the procedures under Self Tests, and
those under Functional Tests. (See Functional tests on page 103.)
A signal-path compensation must have been done within the recommended
calibration interval and at a temperature within ±5 ºC (±9 ºF) of the present
operating temperature. (If at the time you did the prerequisite Self Tests, the
temperature was within the limits just stated, consider this prerequisite met).
A signal-path compensation must have been done at an ambient humidity
within 25% of the current ambient humidity and after having been at that
humidity for at least 4 hours.
The instrument must have been last adjusted at an ambient temperature
between +18 ºC (+64 ºF) and +28 ºC (+82 ºF), must have been operating for a
warm-up period of at least 20 minutes (60 minutes for the MSO/
DPO73304DX instruments if the humidity is >60%), and must be operating
at an ambient temperature as listed in the specifications. (See Environmental
specifications on page 76.) (The warm-up requirement is usually met in the
course of meeting the Self Tests and Functional Tests prerequisites listed
above).
Support sensor, probe, and adapter setups to avoid stress or torque when
connected to the device under test (DUT).
Equipment required
The performance verification procedures use external, traceable signal sources to
directly check warranted characteristics. The following table lists the required
equipment.
1. Attenuator, 10X (two required)
Minimum requirements Example Purpose
Ratio: 10X; impedance 50 Ω;
connectors: female input,
male output
BNC ≤ 2 GHz Tektronix part
number 011-0059-03
SMA ≤ 18 GHz Tektronix part
number 015-1003-00
Signal attenuation, bandwidth,
trigger sensitivity
2. Attenuator, 5X
Ratio: 5X; impedance 50 Ω;
connectors: female input,
male output
BNC ≤ 2 GHz Tektronix part
number 011-0060-02
SMA ≤ 18 GHz Tektronix part
number 015-1002-01
Signal attenuation, bandwidth,
trigger sensitivity
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
116 MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series

Table of Contents

Other manuals for Tektronix DPO5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO5000 Series and is the answer not in the manual?

Tektronix DPO5000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals