Performance Verification
Test Record
Print out the following test record pages and use them to record the performance
test results for your signal analyzer.
RSA5100B Series Test
Record
Model:
Certificate Number:
Technician:
Serial Number:
Calibration Date:
Frequency Accuracy Low limit Test Result High limit
Reference output frequency accuracy
(Standard)
9,999,960 Hz 10,000,040 Hz
(Option PFR)
9,999,996 Hz 10,000,004 Hz
Reference output power level
0dBm
External reference input level
Fail Pass
Phase Noise (with Option 11) Test Result High limit
Offset
CF + 1 kHz –103 dBc/Hz
CF + 10 kHz –109 dBc/Hz
CF + 100 kHz –112 dBc/Hz
CF + 1 MHz –130 dBc/Hz
CF + 6 MHz –137 dBc/Hz
CF + 10 MHz –137 dBc/Hz
Phase Noise (without Option 11) Cursor Carrier power Phase noise
1
High limit
Offset
CF + 1 kHz –103 dBc/Hz
CF + 10 kHz –109 dBc/Hz
CF + 100 kHz –112 dBc/Hz
CF + 1 MHz –130 dBc/Hz
CF + 6 MHz –137 dBc/Hz
CF + 10 M Hz –137 dBc/Hz
1
Phase noise = –(cursor measurement – carrier power)
134 RSA5100B Series Technical Reference