Specifications
3
Option B16x, B1
6xHD only.
4
Option B125, B125HD only.
Table 16: Channel response after cal/norm (center frequency 65 MHz) (RSA5115B/5126B)
Characteristic Description
Amplitude Flatness
Specification
Span 300 kHz
±0.5 dB
1
1
Checked by t he IF t est at 65 MHz combined with the RF flatness.
Table 17: Pulse measurements, typical
Characteristic Description
85, 125, and 160 MHz BW 25 and 40 MHz BW
Minimum Pulse Width for detection, typical
50 ns 150 ns
±0.3 dB + a
bsolute Amplitude Accuracy
Average ON Power
(18°Cto28°C),typical
For pulse widths 100 ns, duty cycles
of 0.5 to
0.001, and S/N ratio 30 dB
For pulse widths 300 ns, and signal
levels >
70 dB below Ref Level
±0.2% o
f reading
Duty Factor, typical
For pulse widths 150 ns, duty cycles
of 0.5 t
o 0.001, and S/N ratio 30 dB
For pulse widths 450 ns, duty cycles
of 0.5 t
o 0.001, and S/N ratio 30 dB
±0.4 dB + absolute Amplitude AccuracyAverage Transmitted Power, typical
For pulse widths 100 ns, duty cycles
of 0.5 to 0.001, and S/N ratio 30 dB
For pulse widths 300 ns, duty cycles
of 0.5 to 0.001, and S/N ratio 30 dB
±0.4 dB + absolute Amplitude AccuracyPeak Pulse Power, typical
For p
ulse widths 100 ns, duty cycles
of 0.5 to 0.001, and S/N ratio 30 dB
For p
ulse widths 300 ns, duty cycles
of 0.5 to 0.001, and S/N ratio 30 dB
±0.25% of reading
Puls
eWidth,typical
For
pulse widths 150 ns, duty cycles
of 0.5 to 0.001, and signal levels >50 dB
below Ref Level
For
pulse widths 450 ns, duty cycles
of 0.5 to 0.001, and S/N ratio 30 dB
System Rise time, typical <12 ns (85, 125 MHz BW)
<7
ns (160 MHz BW)
<40 ns (25 MHz BW)
<2
5ns(40MHzBW)
RSA5100B Series Technical Reference 17