Specifications
Performance C
onditions
The performance limits in these specifications are valid with these conditions:
The signal analyzer must have been calibrated and adjusted at an ambient
temperature between +18 °C and +28 °C.
The signal analyzer must be in an environment with temperature, altitude,
humidity, and vibration within the operating limits described in these
specifications.
Ambient temperature must remain stable while executing the self-alignment.
The signal analyzer must have had a warm-upperiodofatleast20minutes
after starting the RSA5100B application.
Electri
cal S pecifications
Table 2: Frequency
Charac
teristic
Descri
ption
Measur
ement frequency
LF band 1 Hz to 32 MHz
RSA5103B (RF band) 9 kHz to 3 GHz
RSA5106B (RF band) 9 kHz to 6.2 GHz
RSA5115B (RF band) 1 MHz to 15 GHz
Frequency
range
, nominal
RSA5126B (RF band) 1 MHz to 26.5 GHz
Readout Accuracy
±(RE × M F + 0.001 × S pan + 2) Hz
RE
: Reference Frequency Error
MF: Marker Frequency [Hz]
Frequency
Marker
Re
adout Resolution
Reference level dependent
As small as 0.0001 µV
Residual FM, typical
<2 Hz
p-p
in 1 second at 200 MHz CF, 100 Hz span, Freq vs Time
m
ode, Autoscale (95% confidence)
Span Accuracy ±0.3% of span (Auto mode)