Brief Procedures
1–6
TDS 500D, TDS 600C, TDS 700D & TDS 714L Performance Verification and Specifications
H Press SHIFT; then press UTILITY.
H Highlight Cal in the pop-up menu by repeatedly pressing the
main-menu button System. See Figure 1–2.
H Verify that the word Pass appears in the main menu under the
following menu labels: Voltage Reference, Frequency Response,
and Pulse Trigger. See Figure 1–2.
f. Run the signal-path compensation: Press the main-menu button Signal
Path; then press the side-menu button OK Compensate Signal Paths.
g. Wait: Signal-path compensation may take five to fifteen minutes to run.
While it progresses, a “clock” icon (shown at left) is displayed on-
screen. When compensation completes, the status message will be
updated to Pass or Fail in the main menu. See step h.
h. Confirm signal-path compensation returns passed status: Verify that the
word Pass appears under Signal Path in the main menu. See Figure 1–2.
Verify Pass
3
Highlight Cal
1
Verify Pass
2
Figure 1–2: Verifying adjustments and signal-path compensation
2. Return to regular service: Press CLEAR MENU to exit the system menus.