Performance Tests
TDS 500D, TDS 600C, TDS 700D & TDS 714L Performance Verification and Specifications
173
Oscilloscope
50 W coaxial cable
Female BNC to
clip adapter
Black lead
to GND
Figure 121: Initial test hookup
b. Initialize the oscilloscope:
H Press save/recall SETUP.
H Press the main-menu button Recall Factory Setup.
H Press the side-menu button OK Confirm Factory Init.
c. Modify the initialized front-panel control settings:
H Set the horizontal SCALE to 200 ms.
H Press SET LEVEL TO 50%.
H Use the vertical POSITION knob to center the display on screen.
H Press SHIFT; then press ACQUIRE MENU.
H Press the main-menu button Mode; then press the side-menu button
Average.
H Select 128 averages with the keypad or the general purpose knob.
2. Confirm that the Probe Compensator signal is within limits for frequency:
a. Measure the frequency of the probe compensation signal:
H Press MEASURE; then press the main-menu button Select
Measrmnt for Ch1.
H Repeatedly press the side-menu button more until Frequency
appears in the side menu (its icon is shown at the left). Press the
side-menu button Frequency.
b. Check against limits:
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Tektronix TDS 700D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 700D and is the answer not in the manual?

Tektronix TDS 700D Specifications

General IconGeneral
BrandTektronix
ModelTDS 700D
CategoryTest Equipment
LanguageEnglish

Related product manuals