Brief Procedures
TDS 500D, TDS 600C, TDS 700D & TDS 714L Performance Verification and Specifications
115
H Turn the general purpose knob to select the file to recall. For example, if
you followed the instructions previously and saved the file to either the
floppy disk drive (fd0:), or the optional hard disk drive (hd0:), you had
the oscilloscope assign the name TEK00000.SET to your file.
H Press the side-menu button Recall From Selected File.
H Verify that the oscilloscope retrieved the saved setup from the disk. Do
this by noticing the horizontal SCALE for the M (main) time base is
again 200 ms and the waveform shows only two cycles just as it was
when you saved the setup.
3. Remove the test hookup:
H Disconnect the probe from the channel input and the probe-compensa-
tion terminals.
H Remove the floppy disk from the floppy disk drive.