Performance Tests
1–54
TDS 500D, TDS 600C, TDS 700D & TDS 714L Performance Verification and Specifications
H Press VERTICAL MENU; then press the main-menu button
Coupling. Press the side-menu button W to toggle it to the 50 W
setting. (This step is not required on theTDS694C or TDS794D.)
H Press SET LEVEL TO 50%.
H Use the vertical POSITION knob to center the test signal on screen.
H Press TRIGGER MENU; then press the main-menu button Mode
& Holdoff. Press the side-menu button Normal.
2. Confirm Main and Delayed time bases are within limits for accuracies:
a. Display the test signal:
H Press HORIZONTAL MENU.
H Set horizontal modes. To do this, press the main-menu button Time
Base. Press the side-menu buttons Delayed Only and Delayed Runs
After Main. See Figure 1–13 on page 1–55.
b. Measure the test signal:
H Set the horizontal SCALE of the D (delayed) time base to 500 ns for
the TDS600C or to 100 ns for the TDS694C, TDS500D, TDS700D,
or TDS714L.
H If using a time-mark generator, align the trigger T to the center
vertical graticule line by adjusting the horizontal POSITION. See
Figure 1–13 on page 1–55.
H If using a sine wave generator, align the rising edge of the sine wave
on the center graticule crosshairs by adjusting the horizontal
POSITION.
H Set delayed time to 10.000016 ms. Do this on the keypad by
pressing 10.000016, then SHIFT, then m followed by ENTER).
c. Check long-term sample rate and delay time accuracies against limits:
H CHECK that the rising edge of the marker (or sine wave) crosses
the center horizontal graticule line at a point within either ±1.0
divisions for the TDS694C, ±2.0 divisions for the TDS600C,
±2.5 divisions for the TDS500D, TDS700D, or TDS714L, of center
graticule.
See Figure 1–13 on page 1–55.
H Enter number of divisions on the test record.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com