Triggering on Waveforms
3–78
TDS 500D, TDS 600B, & TDS 700D User Manual
The slope control determines whether the oscilloscope finds the trigger point on
the rising or the falling edge of a signal. (See Figure 3–38.)
You set trigger slope by first selecting Slope in the Main Trigger menu and then
selecting between the rising or falling slope icons in the side menu that appears.
The level control determines where on that edge the trigger point occurs. (See
Figure 3–38.) The oscilloscope lets you set the main trigger level with the trigger
MAIN LEVEL knob.
Positive-Going Edge Negative-Going Edge
Trigger slope can be positive or negative.
Trigger level
can be adjusted
vertically.
Figure 3–38: Slope and Level Controls Help Define the Trigger
The oscilloscope also has a delayed trigger system that provides an edge trigger
(no pulse or logic triggers). When using the delayed time base, you can also
delay the acquisition of a waveform for a user-specified time or a user-specified
number of delayed trigger events (or both) after a main trigger event. See
Delayed Triggering on page 3–117 to learn how to use delay.
Triggering from the Front Panel
The trigger buttons and knob let you quickly adjust the trigger level or force a
trigger. (See Figure 3–39.) The trigger readout and status screen lets you quickly
determine the state of the trigger system. You use the following trigger controls
and readouts for all trigger types except where noted.
To manually change the trigger level when edge triggering (or certain threshold
levels when logic or pulse triggering), turn the MAIN LEVEL knob. It adjusts
the trigger level (or threshold level) instantaneously no matter what menu, if any,
is displayed.
Slope and Level
Delayed Trigger System
To set MAIN LEVEL

Table of Contents

Other manuals for Tektronix TDS 700D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 700D and is the answer not in the manual?

Tektronix TDS 700D Specifications

General IconGeneral
BrandTektronix
ModelTDS 700D
CategoryTest Equipment
LanguageEnglish

Related product manuals