Acquiring and Displaying Waveforms
3–38
TDS 500D, TDS 600B, & TDS 700D User Manual
Figure 3–17: Acquire Menu — Stop After
Press RUN/STOP button only (side) to use the RUN/STOP button to start
or stop acquiring. Pressing the RUN/STOP button once will stop the
acquisitions. The upper left hand corner in the display will say “Stop” and
show the number of acquisitions. If you press the button again, the oscillo-
scope will resume taking acquisitions.
Press Single Acquisition Sequence (side). That selection lets you run a
single sequence of acquisitions by pressing the RUN/STOP button. In
Sample, Peak Detect, or Hi Res mode, the oscilloscope will acquire a
waveform record with the first valid trigger event and stop. (Hi Res is
available only on TDS 500D and TDS 700D models.)
In Envelope or Average mode, the oscilloscope will make the specified
number of acquisitions to complete the averaging or enveloping task.
TDS 500D and TDS 700D models only: If the oscilloscope is in equivalent-
time mode and you press Single Acquisition Sequence (side), it will
continue to recognize trigger events and acquire samples until the waveform
record is filled.

Table of Contents

Other manuals for Tektronix TDS 700D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 700D and is the answer not in the manual?

Tektronix TDS 700D Specifications

General IconGeneral
BrandTektronix
ModelTDS 700D
CategoryTest Equipment
LanguageEnglish

Related product manuals