Checking Chroma
/Luma Delay (Lightning Display)
Checking Chro
ma/Luma Delay (Lightning Display)
The Lightning display can be used for interchannel timing measurements. When the color-difference signal is not coincident
with luma, the transitions between the color dots will deviate from the center mark of a delay scale. The amount of deviation
represents the relative signal delay between luma and the color-difference signal.
1. Connect a signal containing color bar
information, terminating the signal
properly.
2. Press and hold the VECT button to
display the signal and the Vector pop-up
menu.