Communication with Modbus
A43/A44 132 2CMC48001M0201
User Manual Revision: C
9.7 Demand
Mapping table
Header
Data blocks
Note Before you can use the information in this chapter you must be familiar with
and understand the information and the concepts described in “Historical Data” on
page - 114.
The following table shows an overview of the mapping table:
Function
Details
Start Reg
Size
(Hex)
Demand
Header
8300
16
Demand
Data block 1
8310
115
Demand
Data block 2
8390
115
Demand
Data block 3
8410
115
Demand
Data block 4
8490
115
Demand
Data block 5
8510
115
Demand
Data block 6
8590
115
Demand
Data block 7
8610
115
The following table describes the header:
Function
Start
Size
Description
Read/
Reg
write
(Hex)
Get next entry
8300
1
Write value 1 to this register to load the
R/W
next block of values and timestamp
Entry number
8301
1
Write to this register to choose an entry
R/W
number to start reading from
Date/Time
8304
3
Write to this register to choose a date/time
R/W
to start reading from
Direction
8307
1
Write to this register to choose the
R/W
direction of reading
The Data blocks contain the history of demand. Data block 1 to 7 have the
same structure. Each block can contain up to 8 channels. Consequently, in a
meter with 50 demand channels, there are 8 channels in each of block 1 to
block 6 and 2 channels in block 7.
The registers of unused channels are filled with 0xFFFF.