CoriolisMaster FCB330, FCB350, FCH330, FCH350 | CI/FCB300/FCH300-EN Rev. F 37
7.2 Flowmeter sensor model FCB3xx / FCH3xx
7.2.1 Temperature class
Model FCx3xx-A1Y… or FCx3xx-S1Y… Zone 1
Ambient temperature ≤40 °C (≤104 °F) ≤50 °C (≤122 °F) ≤60 °C (≤140 °F)
T1 200 °C (392 °F) 200 °C (392 °F) 200 °C (392 °F)
T2 200 °C (392 °F) 200 °C (392 °F) 200 °C (392 °F)
T5 85 °C (185 °F) 85 °C (185 °F) 75 °C (167 °F)
T6 65 °C (149 °F) 65 °C (149 °F) 60 °C (140 °F)
Model FCx3xx-A2Y… or FCx3xx-S2Y… Zone 2
Temperature class
T1 200 °C (392 °F) 200 °C (392 °F) 180 °C (356 °F)
T2 200 °C (392 °F) 200 °C (392 °F) 180 °C (356 °F)
T4 115 °C (239 °F) 115 °C (239 °F) 115 °C (239 °F)
T6 60 °C (140 °F) 60 °C (140 °F) 60 °C (140 °F)
Ambient and process conditions:
20 ...
60 °C (-4 ... 140 °F)
-40 ... 60 °C (-40 ... 140 °F) (only for integral mount design devices)
-50 ... 200 °C (-58 ... 392 °F)
IP 65, IP 67, and NEMA 4X