Key Reference 4
N9310A User’s Guide 55
AM Depth Pressing this softkey enters the value of AM depth. The
active entry area displays the current value of AM depth.
The AM Depth ranges from 0 % to 100 %, with minimum
increment of 0.1 %.
Default value: 0.0 %
Key sequence:
> AM Depth
AM Source Pressing this softkey reveals a menu of choices for amplitude
modulation sources. It allows you to select an internal
source or an external source from MOD IN connector, or
select internal and external simultaneously.
Pressing INT connects an internally calibrated signal to the
modulator. This internal AM source defaults to a sine wave
at a frequency of 1.0000 kHz.
Pressing EXT connects an externally calibrated signal to the
modulator. You need also set the EXT Coupling, see “Ext
Coupling” on next page.
Pressing INT+EXT connects both an internally calibrated
signal and an externally calibrated signal to the modulator
simultaneously.
Default value: INT
Key sequence:
> AM Source
AM Rate Pressing this softkey modifies the internal modulation
frequency. The active entry area displays the current value
of AM rate. The allowed value range is 20 Hz to 80 kHz. The
minimum increment is 0.1 Hz. when external AM source is
selected, this softkey is disabled.
Default value: 1 kHz
Key sequence:
> AM Rate
AM
AM
AM

Table of Contents

Other manuals for Agilent Technologies N9310A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N9310A and is the answer not in the manual?

Agilent Technologies N9310A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN9310A
CategoryTest Equipment
LanguageEnglish

Related product manuals