SETUP REVIEW
The Series 200 allows several operational parameters to be reviewed and changed without
breaking the calibration seal. These operational parameters are:
Power Up Zero Reset Enable/Disable
Time Format
Digital Output Control Enable/Disable
Sleep Mode Feature Enable/Disable (
Models 205 & 210 Only
)
Auto Shutoff Feature Enable/Disable (
Models 205 & 210 Only
)
Serial Input / Output Configuration
Baud Rate
Parity
Number of Data Bits
Number of Stop Bits
Continuous Output Port 1
Continuous Output Fomat
Continuous Output Port 2
Continuous Output Fomat
Print Tab Settings
Printer Port Selection
Time
Date
Consectutive Number
Gross Weight
Tare Weight
Net Weight
Gross Weight Accumulator
Net Weight Accumulator
Count (
Model 210 Only
)
Piece Weight (
Model 210 Only
)
To enable the Setup Review feature, with the indicator ON:
1. Press the
ASTERISK
key. The indicator will respond by showing the FunCt= (Function)
prompt and alternately flashing off and on (all ON, weighing unit off, then all OFF, weighing
unit ON) the annunciators.
2. Press the
ZERO/REVIEW
key. The display will change to the prompt for the selection of
power-up zeroing (PUO=).
3. Using the same procedure as described in the Setup and Calibration section of this manual,
make the required changes.
4. Press the ASTERISK key to return to the previous prompt.
5. To exit Setup Review, press the
ENTER
key to step the remaining prompts
OR
at anytime,
cycle the power (press the
ON/OFF
key twice).
CALIBRATION C NUMBERS
The C numbers are displayed only during the Test mode operation by pressing the
ASTERISK
key then the
UNITS/TEST
key. The "C" numbers are shown at the end of the test operation and
each number is displayed for approximately 4 seconds, allowing you to record them. Each number
may be up to three (3) digits in length. By recording these numbers you will be able to return the
indicator to its present calibration settings without using test weights simply by entering the C
numbers.
Refer to the Setup and Calibration, "CAL" section of this manual for instructions on
using the "C" numbers
.
If any components have been changed that affect calibration and/or your scale is
used in a commercial application and must be "Legal for Trade" you can not use
"C" numbers to re-calibrate.

Other manuals for Cardinal 200 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Cardinal 200 Series and is the answer not in the manual?

Cardinal 200 Series Specifications

General IconGeneral
BrandCardinal
Model200 Series
CategoryAccessories
LanguageEnglish

Related product manuals