43
Reference Manual
00809-0100-4108, Rev BA
Section 4: Electrical Installation
March 2014
Electrical installation
The LOI password can be configured and enabled/disabled by HART communication via a field
communicator, AMS Device Manager, or the LOI.
Configuring LOI password with field communicator
Configuring LOI password with AMS Device Manager
1. Right click on the device and select Configure.
2. Under Manual Setup select the Security tab.
3. Within the Local Operator Interface click the Configure Password button and follow the
screen prompts.
Configuring LOI password using local operator interface
Figure 4-3. Local operator interface password
4.5 Setting transmitter alarm
On the electronics board is an alarm switch, reference Figure 4-2 on page 43 for switch location.
Follow the steps below to change the alarm switch location.
1. Set loop to manual and remove power.
2. Remove transmitter housing cover.
3. Use a small screwdriver to slide switch to desired position.
4. Replace transmitter cover; cover must be fully engaged to comply with explosion proof
requirements.
From the HOME screen, enter the Fast Key sequence
Device Dashboard Fast Keys
2, 2, 6, 5, 2
EXTENDED MENU
CALIBRAT
DAMPING
TRANSFER FUNCT
SCALED VARIAB
ASSIGN P
TAG
ALARM SAT VALUES
PASSWORD
PASSWORD
SIMULATE
HART REV
BACK TO MENU
EXIT MENU
PASSWORD
PASSWORD ENABLE
CHANGE PASSWORD
BACK TO MENU
EXIT MENU
VIEW CONFIG
ZERO TRIM
UNITS
RERANGE
LOOP TEST
DISPLAY
EXTENDED MENU
EXTENDED MENU
EXIT MENU