EasyManuals Logo
Home>Erbe>Medical Equipment>APC 3

Erbe APC 3 Service Manual

Erbe APC 3
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #22 background imageLoading...
Page #22 background image
6 Remedying malfunctions
22 / 70
80116-941_V24392
2023-03
ERROR list as at: 04/18
APC/APC-A error
Display on the unit: APC or APC-A, detecting module: GCM
Error code Description Additional information Action
APC 1 CRC check triggers an
alarm.
Restart the unit. If the
error occurs again, notify
Erbe Service.
APC 2 ErrorHandler Buffer Over-
flow.
Restart the unit. If the
error occurs again, notify
Erbe Service.
APC 5 – 8 Restart the unit. If the
error occurs again, notify
Erbe Service.
APC 20 – 38 Restart the unit. If the
error occurs again, notify
Erbe Service.
APC 40 – 46 Restart the unit. If the
error occurs again, notify
Erbe Service.
APC 50 – 59 Restart the unit. If the
error occurs again, notify
Erbe Service.
APC 122 EEPROM:
Read error.
Restart the unit. If the
error occurs again, notify
Erbe Service.
APC 123 EEPROM:
Write error.
Restart the unit. If the
error occurs again, notify
Erbe Service.
APC 128 –
134
If the advisory message
persists and can be repro
-
duced, notify Erbe Service.
APC 150 Internal unit temperature
> 80°C.
Terminate activation and
allow the unit to cool
down.
APC 151 Controller temperature
> 80°C.
Terminate activation and
allow the unit to cool
down.
APC 246 –
255
If the advisory message
persists and can be repro
-
duced, notify Erbe Service.

Other manuals for Erbe APC 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Erbe APC 3 and is the answer not in the manual?

Erbe APC 3 Specifications

General IconGeneral
BrandErbe
ModelAPC 3
CategoryMedical Equipment
LanguageEnglish

Related product manuals