6 • Remedying malfunctions
80116-941_V24392
2023-03
S/S-A error
Display on the unit: S or S-A, detecting module: SCM
Error code Description Additional information Action
S 1 Checksum calculation
application check triggers
an alarm.
Faulty software. Restart the unit. If the
error occurs again, notify
Erbe Service.
S 3 Checksum calculation
mode list triggers an
alarm.
Faulty software. Restart the unit. If the
error occurs again, notify
Erbe Service.
S 4 Checksum calculation ver-
sion list triggers an alarm.
Faulty software. Restart the unit. If the
error occurs again, notify
Erbe Service.
S 10 Mainboard voltage 12 V
not within ±10%.
Restart the unit. If the
error occurs again, notify
Erbe Service.
S 11 Mainboard voltage 5 V not
within ±10%.
Restart the unit. If the
error occurs again, notify
Erbe Service.
S 63 – 74 Restart the unit. If the
error occurs again, notify
Erbe Service.
S 78 – 84 Restart the unit. If the
error occurs again, notify
Erbe Service.
S 88 – 94 Restart the unit. If the
error occurs again, notify
Erbe Service.
S 128 SCM start triggered by
watchdog.
Restart the unit. If the
error occurs again, notify
Erbe Service.
S 148 Rotational speed of the
main fan insufficient.
Restart the unit. If the
error occurs again, notify
Erbe Service.
S 159 Invalid instrument param-
eters. UI does not accept
instrument parameters.
Check the accessories. If
the advisory message
occurs again, notify Erbe
Service.
S-A 18 Heat management: Max.
average power output
reached.
Allow the unit to cool
down. Activation may only
be repeated once the unit
has cooled down.
If the error occurs again,
notify Erbe Service.