7 Maintenance and servicing
38 / 70
80116-941_V24392
2023-03
Performance tests
Test setup
The test specimen is connected via the VIO 3 electrosurgical unit to the
power supply.
Test procedure
Socket lighting 1. Call up the "Menu" screen.
2. Call up <Service>.
3. Enter the password VIO3 and confirm with <Entry>. The unit shows
all service settings.
4. Call up <Test programs>. The unit shows all test programs.
5. Call up <Test Program Socket Lightning>.
6. Press <all lights on>. All sockets must be uniformly lit up in white.
7. Then press <green>. All sockets must be uniformly lit up in green.
Proceed in the same way for the colors blue, yellow and red.
8. Terminate the test program with <Close>.
Socket configuration 1. Call up <Test Program Plugging recognition>.
2. Compare the socket configuration on the unit with the socket config-
uration shown in the test program.
The name of the socket in the test program (TP) can deviate here as
follows from the name of the socket on the unit:
TP: MF-O = unit: MF (multifunction socket)
TP: MF-U = unit: MF-U (multifunction socket, universal)
TP: ME = unit: MO (monopolar socket)
TP: BE = unit: BI (bipolar socket)
TP: NE = unit: NE (return electrode socket)
TP: APC = APC unit (APC socket)
3. Terminate the test program with <Close>.
4. Exit the Service settings.