7 • Maintenance and servicing
80116-941_V24392
2023-03
Gas flow measurement
Testing and measuring equipment
Test setup
Fig. 7-5
IMPORTANT! For this test only Erbe pressure reducers may be used. For article num-
bers see chapter "Testing and measuring equipment".
Erbe Art. No. Description
– Flow meter (at least 10 standard liters, 0 °C,
1013
mbar, Luer-Lock)
20100-165 APC test hose
20134-XXX
1
1. dependent on bottle connector.
Pressure regulator
– Argon gas bottle
20100-175 FiAPC test adapter
20194-070 Patient cable NE
or
20194-075 Patient cable NE, international
20100-033 Adapter cable NE
20189-350 Dual-pedal footswitch with ReMode