51 / 70
7 • Maintenance and servicing
80116-941_V24392
2023-03
4. Use the COAG pedal on the footswitch to activate the test specimen.
5. Record the measured value.
6. Repeat the test using all the target flow values in the table below. All
measured values must be within the specified tolerance ranges.
Testing the APC socket
Instrument recognition
via EEPROM
Testing and measuring equipment
Test setup
• The test specimen is connected via the VIO 3 electrosurgical unit to the
power supply.
Test procedure
1. Call up <Test Program Instrument recognition>.
2. Insert the APC applicator into the APC socket. The test program must
show the applicator data record on APC 3, slot 2.
3. Terminate the test program with <Close>.
4. Stay in the test program list for the next test step and leave the appli-
cator inserted in the test specimen.
Targ et fl o w va lu e Tolerance range
0.5 l/min 0.40 – 0.60 l/min
1.0 l/min 0.80 – 1.20 l/min
2.0 l/min 1.60 – 2.40 l/min
5.0 l/min 4.00 – 6.00 l/min
Erbe Art. No. Description
20132-250 APCapplicator