GE HEALTHCARE PROPRIETARY TO GE
D
IRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 3-7 - Power On/Off 3 - 23
3-7-6 Sleep Mode (R1 to R4)
NOTE: This procedure assumes the system is already up and running.
This feature is not a substitute of a regular shutdown. The system should be shutdown completely at
least once a day to prevent performance issues.
1.) To going into the sleep mode, enter the scan screen and lightly press the Power On/Off switch
once. The System-Exit window is displayed.
2.) Select Sleep. The system performs sequence to go to SLEEP mode.
3.) Unplug the power cable from the wall outlet.
4.) Plug the power cable to the wall outlet.
5.) To exit out of Sleep Mode, press the On/Off switch. Boot time is about 90 seconds.
CAUTION
!! CAUTION:
You need to wait at least one minute after the monitor blacks out before unplugging the
power cable. The system is still in the process of going into Sleep Mode after the monitor
blacks out.