EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ S8 Vet

GE LOGIQ S8 Vet User Manual

GE LOGIQ S8 Vet
418 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #68 background imageLoading...
Page #68 background image
GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
2 - 4 Section 2-2 - General console requirements
2-2-3 EMI Limitations
Ultrasound systems are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transients in the air or wiring. Ultrasound systems also generate EMI. The LOGIQ
S8 complies with limits as stated on the EMC label. However, there is no guarantee that interference
will not occur in a particular installation.
Possible EMI sources should be identified before the unit is installed.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect. These
sources include:
Medical lasers
Scanners
Cauterizing guns
•Computers
•Monitors
•Fans
Gel warmers
Microwave oven
Light dimmers
Mobile phones
In-house wireless phones (DECT phones)
Wireless computer keyboard and mouse
Air conditioning system
High frequency (HF) surgery equipment
General AC/DC adapters
The presence of a broadcast station or broadcast van may also cause interference.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S8 Vet and is the answer not in the manual?

GE LOGIQ S8 Vet Specifications

General IconGeneral
BrandGE
ModelLOGIQ S8 Vet
CategoryMedical Equipment
LanguageEnglish

Related product manuals