EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ S8 Vet

GE LOGIQ S8 Vet User Manual

GE LOGIQ S8 Vet
418 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #154 background imageLoading...
Page #154 background image
GE HEALTHCARE PROPRIETARY TO GE
D
IRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3 - 76 Section 3-12 - Configuring Connectivity
3-12-6 Device
To add a new device,
1.) Press Add.
2.) Type the device name in the Name field.
3.) Type the device's IP address in the IP Address field.
Figure 3-48 Connectivity Device Preset Menu
To ping a device,
1.) Select the device.
2.) Press Ping. If the smiley face smiles, then the connection has been confirmed. If the smiley face
frowns, then the connection has not been made. Check the device name and IP address.
Table 3-11 Device
Preset Parameter Description
Add/Remove Press Add to add a new device; press Remove to delete a device.
Ping Press Ping to confirm that a device is connected.
Properties: Name Type the name of the device.
Properties: IP Address Type the device's IP address.
Properties: AE Title
AE Title of the LOGIQ S8. NOTE: Only available for MyComputer.
Properties: Port Number IP Port Number Used for DICOM, set by default to 104. NOTE: Only available for
MyComputer.
Properties: MAC
Address
Unique network card address. NOTE: Only available for MyComputer.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S8 Vet and is the answer not in the manual?

GE LOGIQ S8 Vet Specifications

General IconGeneral
BrandGE
ModelLOGIQ S8 Vet
CategoryMedical Equipment
LanguageEnglish

Related product manuals