DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 7-7 - Troubleshooting Trees, Instructions and Tech Tips 7 - 81
2.) FibroScan Report and Probe button do not work properly
If FibroScan measurement and another function are performed at the same time, FibroScan
functions occasionally do not work properly.
Example 1: If the button on the FibroScan probe is pressed while FibroScan Report is being
generated, the button may not respond.
Example 2: Do not press the button on FibroScan probe, when FibroScan Report is selected and
the system starts generating the report.
<WORKAROUND>
Do not press the button on FibroScan probe, when FibroScan Report is selected and the system
starts generating the report.
3.) Error message: “FibroScan is not available. A FibroScan module is required”
The following error message may appear when starting FibroScan mode.
<RECOVERY>
Reboot the system.
7-7-17 Fibroscan Probe
Indicator of Fibroscan probe is blinking when select the probe in Fibroscan mode. It’s indicated the
probe is active.