GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 7-7 - Troubleshooting Trees, Instructions and Tech Tips 7 - 77
7-7-12 Mode encoder works twice when pushed button once
The cause may be improper setting or dust in a cap. Before replacing an encoder, verify the setting/dust
of a cap.
NOTE: It occasionally sounds like a correct setting from a cap when assembled it improperly.
7-7-13 Sticky buttons
The cause may be dust/liquid into a button. Call GE Service.
7-7-14 Mode button failure
Other function works when push a mode button / Mode button fail to respond even if you push a button
Call GE Service.