IRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
4 - 22 Section 4-4 - Functional Checks
4-4-5 Performance Tests
The use of test phantoms is only recommended if required by your facility's (customer’s) QA program.
4-4-5-1 Recommended Test Phantoms (optional)
GE Healthcare recommends the RMI 430GS phantom (optional), but it is not required. It is the most
current phantom (optional) recommended to our field service personnel and provides the necessary
targets and extended life necessary for consistent system testing.
Figure 4-19 Performance tests
2
4321
4
6
8
10
12
14
16
RMI 403GS
Lateral Distance
Measurement
Functiona
l
Resolutio
n
Contrast
Resolution
&
Gray Scale
P
hotography
Lateral
Resolution
Axial
Resolution
Axial Distance
Measurement
Penetration