GE HEALTHCARE PROPRIETARY TO GE
D
IRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 3-12 - Configuring Connectivity 3 - 75
3-12-5 TCPIP (cont’d)
NOTE: Reboot the system to activate any changes saved from this page.
Table 3-10 IP settings
Preset Parameter Description
Enable DHCP DO NOT select this box to enable dynamic IP Address selection.
NOTE: The system shall disable IP-Address, Subnet Mask, and Default Gateway
when the user chooses to use DHCP.
IP-Address Type the IP Address of the Ultrasound system. NOTE: IP stands for Internet
Protocol. Every device on the network has a unique IP address.
Subnet Mask Type the subnet mask address. NOTE: The Subnet Mask is an IP address filter that
eliminates communication/messages from network devices of no interest to your
system.
Default Gateway Type the default gateway address.
Network Speed Select the network speed (Auto Detect, 10Mbps/Half/Full Duplex, or 100 Mbps/Half/
Full Duplex/1000Mbps/Auto-negotiate)