DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
2 - 10 Section 2-3 - Facility needs
2-3-6 Networking setup requirements
For detail of Connectivity setup information, refer to the Basic User Manual supplied with the each
system.
2-3-6-1 Stand alone scanner (without network connection)
None.
2-3-6-2 Scanner connected to hospital’s network
Supported networks:
Ethernet network connection (R1.x.x/R2 and later)
Up to 900 Mbps WLAN (R2 and later, optional)
10/100/1000 Mbps Ethernet/DICOM network connection is required.
2-3-6-3 InSite Requirements
Need internet access available to be able to connect to Insite.
Insite requires an Ethernet connection with a 10/100 Mbps or 10/100/1000 Mbps interface.
2-3-6-4 Purpose of the DICOM network function
DICOM services provide the operator with clinically useful features for moving images and patient
information over a hospital network.
Examples of DICOM services include the transfer of images to workstations for viewing or transferring
images to remote printers.
As an added benefit, transferring images in this manner frees up the on-board LCD and peripherals,
enabling viewing to be done while scanning continues.
With DICOM, images can be archived, stored, and retrieved faster, easier, and at a lower cost.
2-3-6-5 DICOM Option Pre-installation Requirements
To configure the LOGIQ S8 to work with other network connections, the site’s network administrator
must provide information to complete the form in Figure 2-2 Worksheet for DICOM Network Information
on page 2-11. Ensure that there are no spaces in any field of the form.
Entries must include:
A host name, local port number, AE Title, IP address and Net Mask for the LOGIQ S8.
The IP addresses for the default gateway and other routers at the site for ROUTING
INFORMATION.
The host name, IP address, port and AE Title for each device the site wants connected to the
LOGIQ S8 for DICOM APPLICATION INFORMATION. A field for the make (manufacturer) and the
revision of the device, is also included. This information may be useful for error solving.

Table of Contents

Other manuals for GE LOGIQ S8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S8 and is the answer not in the manual?

GE LOGIQ S8 Specifications

General IconGeneral
BrandGE
ModelLOGIQ S8
CategoryMedical Equipment
LanguageEnglish

Related product manuals