GE HEALTHCARE PROPRIETARY TO GE
D
IRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 3-6 - Completing the Set Up 3 - 13
3-6-2 Electrical specifications
3-6-2-1 Verification of the LOGIQ S8’s voltage setting
Verify that the mains voltage specified for the unit is available on-site.
The voltage setting for the unit is found on a label on the back of the system on lower rear frame of the
LOGIQ S8.
3-6-2-2 Electrical Specifications
In the table below, the electrical specifications for LOGIQ S8 include LCD and on board peripherals.
Power Consumption nominal 900 VA including all on-board peripherals.
WARNINGWARNING
CONNECTING A LOGIQ S8 TO THE WRONG VOLTAGE LEVEL WILL MOST LIKELY
DESTROY THE UNIT.
Table 3-10 Electrical Specifications
Model Voltage Tolerances Current Frequency
LOGIQ S8
100 - 120 VAC ±10% 9.0 ... 7.0 A 50/60 Hz (±2%)
220 - 240 VAC ±10% 4.0 ... 3.7 A 50/60 Hz (±2%)