IRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
4 - 38 Section 4-4 - Functional Checks
4-4-16 ECG Check
4-4-16-1 Introduction
The ECG capability on this unit, is intended as use as a trigger for measurements, but can also be
viewed on the screen.
4-4-16-2 Parts needed
•ECG Harness
• ECG Pads, (3 pc)
or
• ECG Simulator
4-4-16-3 Preparations
None
4-4-16-4 ECG Checks
1.) Select Model on the touch panel.
2.) Select Cardiac as application and select Scan.
3.) If you do not see the ECG line, select the ECG tab and press ECG DISPLAY. You should then see
the line.
NOTE: If the ECG tab is not displayed, select Utility ->Application and Check “Show ECG Tab”.
NOTE: Without the ECG option, the ECG touch panel is not displayed.
NOTE: For more information about ECG, refer to the latest revision of the LOGIQ™ S8 Basic User
Manual, Chapter 13.
Plug in the ECG cables with nothing attached, so you pick up noise, the line should display noise.
Table 4-3 ECG Checks
Step Task Expected Result(s)
1. Connect the ECG harness to the connector on
the front of the system
The unit displays a straight curve along the
bottom edge of the image sector on the screen.
2. Connect the three leads to a ECG simulator,
or
Fasten the three ECG Pads to your body and
connect the three leads to respective ECG Pad
When connecting, the signal on the screen will
be noisy
When the connection is completed, a typical
clean ECG signal is displayed.