IRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3 - 90 Section 3-13 - Setting up InSite Connection (R1 to R4)
3-13-8 Configuring Request for Service (RFS)
To configure the LOGIQ S8 to enable the customer to submit an RFS,
1.) Position the Windows pointer on the GE InSite icon at the bottom of the display.
2.) Press the right Trackball Set key. Select Contact GE. The RFS screen opens.
3.) Select the Users tab. Click on Add Users and type the site’s Contact Name. Manual and machine-
generated RFS requests are directed to this person. Items with red asterisks are required fields and
must be filled in.
4.) Click on Set Default Machine contact.
Figure 3-56 Users tab
NOTE: Machine-generated RFS is the system default. If you wish to disable RFS, remove the checkmark at
Automatic Request for Service via Utility -> Admin -> System Admin.
Figure 3-57 System Admin page
3-13-8-1 Remote Check and Configuration
Contact the OnLine Center for InSite Checkout confirmation.