6 - 2 Section 6-2 - Monitor Adjustment
Section 6-2
Monitor Adjustment
6-2-1 19” LCD Monitor
The digital control panel is located at the front of the color monitor. It is NOT recommended to change
the pre-adjusted settings.
However, if you are not satisfied with the factory settings, use these controls to program those you prefer
in each resolution.
NOTE: All changed values will only be saved by selecting “Exit” from the OSD. If not, the adjusted values will
be lost after loss of power.
Figure 6-1 Monitor Adjustment buttons
NOTE: 1280 x 1024, 60Hz.
Table 6-1 Contents in this section
Section
Number
Main Menu Sub Menu Range Setting for LOGIQ S8
6-2-1-1 Picture
Contrast 0~100% 100%
Brightness 0~100% 80%
Colortemp/Gamma
Select
2.2 or 2.4 2.4
Colortemp/Mode
9000K/11000K/13000K/15000K/
USER
13000K
6-2-1-2 Function
Scale Full/5:4/Native Full
Information
Memory Recall Factory Default
SBC ON/OFF ON
6-2-1-3 OSD
Language
English/German/French/Spanish/
Italian/Swedish/Chinese/Japanese
English
H-Position 0~100% 50%
V-Position 0~100% 50%
Half Tone 0~100% 50%
- Exit
When finishing the Adjusting Menu, select the EXIT (middle) and press the MENU
(middle) key.