GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
6 - 6 Section 6-2 - Monitor Adjustment
6-2-1-2-3 Memory Recall
1.) Press the MENU (middle) key of the monitor controls over 10 sec.
2.) Select the FUNCTION -> MEMORY RECALL by pressing the < LEFT or RIGHT > button to move
the cursor and the MENU (middle) key.
NOTE: MEMORY RECALL is Factory default. If selecting MEMORY RECALL, all settings will be back
to factory default status.
6-2-1-2-4 SBC
1.) Press the MENU (middle) key of the monitor controls over 10 sec.
2.) Select the FUNCTION -> SBC by pressing the < LEFT or RIGHT > button to move the cursor and
the MENU (middle) key.
3.) Select ON/OFF by pressing the < LEFT or RIGHT > button to decrease/increase the value and the
MENU (middle) key. (Default : ON)