Connectivity setup
LOGIQ V2/LOGIQ V1 – Basic Service Manual 3-39
5610739-100 English Rev.10
How to Setup and Use a DICOM Image Storage Service
An Image Storage Service provides a place to store patient and
exam data from LOGIQ V2/LOGIQ V1 and corresponding
images. The Image Storage Service, or the device that hosts it,
is often called a Patient Archiving and Communication System
(PACS).
Table 3-14: Setup an Image Storage Service
1In Utility -> Connectivity -> Service, from the Destination Device drop-down menu,
select the device on which the service resides. (This assumes you have already setup
the device in the Device tab in
‘How to Setup and Use a DICOM Image Storage Service’ on
page 3-39
.
2 From the Select Service Type to add drop-down menu, select DICOM Image Storage
and press Add.
3 Change the name of the service to one of your choosing.
4 Enter the AE Title and port Number of the service. AE Title is case-sensitive.
4 Save your changes and then press Verify. A “Smiley Face” indicates successful
communication with the service.
Note: If you get a successful Ping (Smiley Face) at the device level but not at the
service level, it is possible that the AE Title or Port Number of the service settings are
incorrect. Ensure that these are correct then re-verify. Be sure that the service type
(Store, Print, etc) is correct and supported by the device.

Table of Contents

Other manuals for GE LOGIQ V1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V1 and is the answer not in the manual?

GE LOGIQ V1 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V1
CategoryMedical Equipment
LanguageEnglish

Related product manuals